Analysis of the structure and mechanism of destruction of implants manufactured by the additive method. / Koemets, Y.; Kuznetsov, V.; Kazantseva, N. и др.
в: AIP Conference Proceedings, Том 2899, № 1, 020073, 2023.

Результаты исследований: Вклад в журналМатериалы конференцииРецензирование

Harvard

APA

Vancouver

Author

BibTeX

@article{14c3271ab0804438806ebd06c8beb40d,
title = "Analysis of the structure and mechanism of destruction of implants manufactured by the additive method",
abstract = "Studies of the original implant for restoring the function of the tubular bone stump were carried out. The implant was manufactured by the Electron Beam Melting method from 316L steel powder used for medical purposes. A high roughness of the implant surface was found, which is necessary for osseointegration in the bone. Technological defects of the product are revealed, which affect the subsequent destruction and reduce the service life of the product. {\textcopyright} 2023 American Institute of Physics Inc.. All rights reserved.",
author = "Y. Koemets and V. Kuznetsov and N. Kazantseva and M. Karabanalov and O. Koemets and А. Yemanov",
note = "Support by the Russian Science Foundation (project No. 22-29-01514) is acknowledged. The structural studies were done in the collective use center of the Ural Federal University named after First President of Russia B.N. Yeltsin.",
year = "2023",
doi = "10.1063/5.0163737",
language = "English",
volume = "2899",
journal = "AIP Conference Proceedings",
issn = "0094-243X",
publisher = "American Institute of Physics Publising LLC",
number = "1",

}

RIS

TY - JOUR

T1 - Analysis of the structure and mechanism of destruction of implants manufactured by the additive method

AU - Koemets, Y.

AU - Kuznetsov, V.

AU - Kazantseva, N.

AU - Karabanalov, M.

AU - Koemets, O.

AU - Yemanov, А.

N1 - Support by the Russian Science Foundation (project No. 22-29-01514) is acknowledged. The structural studies were done in the collective use center of the Ural Federal University named after First President of Russia B.N. Yeltsin.

PY - 2023

Y1 - 2023

N2 - Studies of the original implant for restoring the function of the tubular bone stump were carried out. The implant was manufactured by the Electron Beam Melting method from 316L steel powder used for medical purposes. A high roughness of the implant surface was found, which is necessary for osseointegration in the bone. Technological defects of the product are revealed, which affect the subsequent destruction and reduce the service life of the product. © 2023 American Institute of Physics Inc.. All rights reserved.

AB - Studies of the original implant for restoring the function of the tubular bone stump were carried out. The implant was manufactured by the Electron Beam Melting method from 316L steel powder used for medical purposes. A high roughness of the implant surface was found, which is necessary for osseointegration in the bone. Technological defects of the product are revealed, which affect the subsequent destruction and reduce the service life of the product. © 2023 American Institute of Physics Inc.. All rights reserved.

UR - http://www.scopus.com/inward/record.url?partnerID=8YFLogxK&scp=85176930885

U2 - 10.1063/5.0163737

DO - 10.1063/5.0163737

M3 - Conference article

VL - 2899

JO - AIP Conference Proceedings

JF - AIP Conference Proceedings

SN - 0094-243X

IS - 1

M1 - 020073

ER -

ID: 48553584